AFS® Water Purification System

Merck

CLRW quality

AFS® systems combine complementary water purification techniques; complete pretreatment, intelligent reverse osmosis (RO), bactericidal UV lamp treatment and a final 0.22 µm filter ensure a reliable and consistent water quality. Water produced by the systems complies with the Clinical and Laboratory Standards Institute® (CLSI) guideline for clinical laboratory reagent water (CLRW). Depending on daily volume needs, different production flow rates are available (8, 16 and 24 L/h).

The systems are designed to fulfill accreditation needs (such as CAP™ 15189 accreditation to the ISO® 15189:2007 standard), and have full monitoring and water quality archiving capabilities that allow storage of up to six months of data. Optional activation of Millitrack® Essential e-Solution allows easy, rapid access to this information and also enhances data management control and provides remote system access.

User friendly & easy maintenance

Robust AFS® systems require little maintenance, which means less analyzer downtime. When maintenance is required, an intuitive icon-based display and a new pack locking system facilitate pack changes. Systems have a small footprint and can be installed wherever it is convenient in the laboratory: on a wall, on or under a bench, or on a cart.

Read more

Model No Model Name Action
Resistivity at 25 °C > 15 MΩ•cm
TOC Typically < 30 ppb
Bacteria < 1 CFU/mL
Silica < 5 µg/L
AFS® 8 8 L/h
AFS® 16 16 L/h
AFS® 24 24 L/h
Feed water quality Potable
Feed water pressure 1 - 6 bar
Feed water temperature 5 - 35 °C
Minimum flow rate feed > 5 L min at 2 bar
Total chlorine < 3 ppm
Langelier Saturation Index < 0.3
Maximum Fouling Index (SDI) < 12
Dimensions (H x W x D) 585 x 268 x 426 mm (23 x 10.6 x 16.8 in)
Operating weight < 28 kg (61.7 lb)
Electric power supply voltage 100 - 240 V +/- 10%
Electric power supply frequency 50 - 60 Hz +/- 10%
AFS® main unit data connection Ethernet (RJ45)

Accessories