flame-photometer-fp8800

Flame Photometers

logo (1)

Krüss offers the Flame Photometer model FP8800 for simple, precise, cost-effective measurement in industrial applications. It is Ideal for concentrated alkaline elements. In aqueous solutions, the unit will deliver a rate of around 300 samples per hour. It measures up to three alkaline elements simultaneously

  • Economic, robust, safe and long-lasting measuring instrument.
  • Simple, precise and cost-efficient measuring method for the simultaneous determination of alkali and alkaline earth element concentrations in aqueous samples.
  • Precision ≤ 0.5 % CV with 10 mg/l Na-, K-, Li-, Ca-standard solution.
  • Touch Screen , 800 x 600 pixels. Complete user management, SQL data base, export of measured data to Excel.
  • Option 1: automatic measuring connected with the sampler and sample dilution (FP8801).
  • Option 2: PROFIBUS DP interface and remote module for usage in process control (FP8800).
  • The FP8800 has 4 independent measuring channels and an additional separate channel for the monitoring of the flame.
  • The 4 measuring channels can be equipped with the appropriate optical filters.
  • The FP8800 does not require a spectral reference line, although measurement operation can be carried out either with or without a reference element.
Parameter Description
Measurement principle Emission flame photometer for the determination of alkali and alkaline earthelement concentrations (Na, K, Li, Ca, …) in aqueous solutions. Others on request
Type of device FP8800 for process analytics, FP8801 for laboratory analytics.Both devices with manual or automatic sample feeding
Total measurement ranges s Na 0,01 – 4500 mg/l  (ppm) ^= 0,0004 – 200 mmol/lK 0,02 – 4500 mg/l      (ppm) ^= 0,0005 – 110 mmol/l

Li 0,02 – 3000 mg/l     (ppm) ^= 0,0028 – 420 mmol/l

Ca 0,30 – 3000 mg/l    (ppm) ^= 0,0075 – 75 mmol/l

 Sensitivity Na < 1 µg/l (ppb)      ^= 0,04 µmol/lK < 2 µg/l (ppb)       ^= 0,05 µmol/l

Li < 2 µg/l (ppb)      ^= 0,28 µmol/l

Ca < 30 µg/l (ppb)  ^= 7,50 µmol/l

Interference filter Na 589 ± 2 nmK 767 ± 2 nm

Ca 622 ± 2 nm

Bandwidth 8 ± 1 nm Multiple filters capture the total measurement range
Characteristic line Linear Linear approximation with 2 standards
Non-linear Cubic approximation with 3 – 8 standards
Reference Calculation of results with or without an internal standardLi-reference solution 35 mg/l (ppm) ^= 5 mmol/l
Samples Volume ~ 2 ml/Probe, through-put ~ 240 samples/h
Drift ≤ 0,1 % after 30 min warming-up time
Precision ≤ 0,2 % CV over 20 results (random error, 1S-variance)with 10 mg/l Na-, K-, Li-, Ca- standard solution
Accuracy ≤1 % of mean value over 20 results (systematical error, recovery)with 10 mg/l Na-, K-, Li-, Ca- standard solution
Methods memory 99 freely definable
Results memory 999 result data sets in a ring buffer
Compressed air Free of oil, water and particlescomsumption ~ 12 l/min, input pressure 1,2 ± 0,3 bar
Gas supply Propan or acetylene for flame photometryconsumption propane ~0,3 l/min , acetylene ~0,5 l/min, input pressure 1,2 ± 0,2 bar
Input, display Touchscreen and USB mouse, 8,4“ (21 cm) TFT-LCD display 800 x 600 Pixel
Operator guidance German, english, other languages on request
Interfaces 1x PROFIBUS DP®     Field bus for FP8800, retrofit for FP88012x USB Mouse,        flash drive for data export and firmware update

2x RS-232                 Printer, sample unit

2x Ethernet              LAN, LIMS

Main voltage 230 V ± 10 %, 50 to 60 Hz ± 5 %, optional 120 V ± 10 %
Power consumption Flame Photometer ≤ 75 W, with compressor ≤ 285 W